Interference of Atrial Fibrillation with Implantable Cardioverter Defibrillator Therapy in Patients with Heart Failure and Reduced Ejection Fraction

    Authors

    Abstract

    Implantable cardioverter defibrillators (ICD) are an important therapeutic option in reducing mortality due to ventricular arrhythmias in patients with heart failure with reduced ejection fraction. Atrial fibrillation (AF) is often present in these patients and may interfere with ICD therapy by inducing inappropriate and appropriate shocks. This issue is important because both types of shock increase mortality in patients with ICD. The strategies to minimize the rate of ICD shocks induced by AF include the optimization of ICD programming, pharmacological therapy for rate or rhythm control, and ablation of AF or atrioventricular junction ablation. In this review, we describe the interference of AF with ICD therapy, examine the impact of this interference on survival, and discuss the strategies for its reduction.

    Keywords

    interferencija, fibrilacija atrija, implantabilni kardioverterski defibrilator, interference, atrial fibrillation, implantable cardioverter defibrillator

    DOI

    https://doi.org/10.15836/ccar2024.177

    Full Text

    ## Introduction Implantable cardioverter defibrillators (ICD) are a well-established and effective therapy for patients with heart failure with reduced ejection fraction (HFrEF) who survived sustained ventricular tachycardia (VT) or ventricular fibrillation (VF), or those who are exposed to an increased risk of these arrhythmias. (1) With regard to poor left ventricular systolic function, AF is often present in patients with ICD, with a prevalence from 17% to 36% at time of implantation in clinical practice. (2, 3) Except for the negative effect on survival, (4, 5) AF may interfere with ICD therapy and limits its clinical usefulness by inducing inappropriate and appropriate shocks. This issue is important because ICD shocks are painful, cause psychological disturbances, and may induce ventricular pro-arrhythmia and even death. (6-9) The aim of this review was to explore the possible interference mechanisms of AF with ICD therapy, examine the impact of this interference on survival, and discuss the treatment options available for its reduction. Mechanisms of atrial fibrillation interference with implantable cardioverter-defibrillator therapy ICDs are an important therapeutic option in reducing mortality due to ventricular arrhythmias in patients with HFrEF. (10) The primary goal of ICD therapy is to detect and stop the life-threatening ventricular arrhythmia by delivering electrical shock and thus prevent a sudden cardiac death. In that context, ICD shocks delivered for VT/VF are considered appropriate because they save lives. (11) In contrast, ICD shocks delivered for the non-ventricular arrhythmias, like AF, atrial flutter (AFL), supraventricular tachycardia (SVT), atrial tachycardia (AT), sinus tachycardia (ST), or non-arrhythmic events, like electromagnetic interference, myopotentials, and device malfunction or oversensing, are unnecessary and are considered inappropriate. (12) AF may interfere with ICD therapy in two ways. The first way is more common, and occurs when the fast ventricular rate of AF with relatively regular R-R intervals reaches a device’s programmed detection zone for VT/VF, is misclassified as ventricular arrhythmia, and induces the delivery of inappropriate shock (**Figure 1**). According to data from randomized clinical trials, a history of AF was found in 9% to 24% of patients with ICD (**Table 1**), more often in patients with non-ischemic than in those with ischemic cardiomyopathy. (13-19) On the other hand, the incidence of new-onset AF lasting at least 24 hours per day was 6.3% and 7.1% in single chamber and doubled chamber patients with ICD, respectively. (20) Therefore, it is not surprising that AF/AFL was the most common reason for an inappropriate ICD shock, followed by ST/SVT and non-arrhythmic events. (21, 22) In earlier randomized clinical trials, (15, 21, 23, 24) the rate of inappropriate ICD shocks ranged from 15% to 21% over 20 to 45.5 months of follow-up (**Table 1**). Later, thanks to device programming adjustment and enhanced algorithm discrimination of arrhythmias, the rate of inappropriate shocks decreased, (18, 19, 25) but even with these improvements, they occur in 1.9% to 3.7% of patients with ICD over a period of two years. (26, 27) FIGURE 1. Inappropriate shock induced by atrial fibrillation in a patient with a double-chamber implantable cardioverter defibrillator. AS = ATRIAL SENSING, VS = VENTRICULAR SENSING, CD = CARDIOVERSION. ### TABLE 1: Summary of atrial fibrillation history and implantable cardioverter-defibrillator shocks in randomized controlled trials. | **RCT** | **No of ICD** **patients** | **Heart disease** **(%)** | **History of AF (%)** | **EF** **(median)** | **FU (m)** **(mean)** | **ICD shocks (%)** **Approp Inapprop** | | --- | --- | --- | --- | --- | --- | --- | | AVID, (13, 23) | 507 | ICM 81 DICM 19 | 21 | 32% | 18 | 39 20 | | MADIT II (14, 21) | 719 | ICM 100 | 9 | 23% | 20 | 18 15 | | DEFINITE (15) | 229 | DICM 100 | 23 | 29% | 29 | 18 21 | | SCD-HeFT (16, 24) | 811 | ICM 53 DICM 47 | 17 | 25% | 45 | 22 17 | | MADIT-CRT (17, 25) | 1790 | ICM 55 DICM 45 | 11 | 28% | 39 | 13 7 | | MADIT-RIT (18) | 1500 | ICM 53 DICM 47 | 13 | 26% | 17 | 13 4 | | DANISH (19) | 556 | DICM 100 | 24 | 25% | 67 | 12 6 | [†] RCT = randomized controlled trial, No = number, AF = atrial fibrillation, EF = ejection fraction, FU = follow-up, m = months, ICD = implantable cardioverter-defibrillator, approp = appropriate, inapprop = inappropriate, ISM = ischemic cardiomyopathy, NICM = non-ischemic cardiomyopathy. The second way in which AF may interfere with ICD therapy is direct action of AF in causing an episode of VT/VF, with consequent appropriate shock therapy. An earlier analysis of stored ICD electrograms revealed that approximately 10% of VT episodes were preceded by atrial fibrillation typically associated with a rapid ventricular response. (28) Subsequently, Gronefeld et al. (29) reported that patients in AF experienced appropriate ICD therapy for recurrent ventricular arrhythmias more frequently than those in sinus rhythm (63% vs. 38%, p=0.01) because of a higher incidence of short-long-short cycles preceding ventricular arrhythmias in AF (50% vs. 16%, p=0.002). In the study by Stein et al. (30) which included 537 patients with dual-chamber ICD, 233 (8.6%) of 2602 VT/VF episodes were preceded by a paroxysm of AT or AF. The median duration of these arrhythmias preceding VT/VF was approximately one hour. Ventricular therapies were delivered to treat 205 of these VT/VF episodes, while the remaining 28 episodes terminated spontaneously before an ICD therapy could be delivered. On the other hand, Borleffs et al. reported that patients with permanent AF had higher risk of ventricular arrhythmias triggering ICD discharge compared with those without AF and those with paroxysmal or persistent AF (49% vs. 29%, 26%, and 26%, respectively; p<0.001). (4) There are several mechanisms that can explain the association between AF and appropriate ICD shocks. First, it is possible that AF and ventricular arrhythmias share risk factors such as ischemia, renal failure, increased sympathetic tone, or increased left ventricular filling pressure, (31, 32) which facilitate the initiation of both arrhythmias. Second, in structural heart disease with reduced EF, the rapid ventricular response caused by AF may directly affect ventricular refractoriness and therefore induce ventricular arrhythmia (**Figure 2**). (30, 33) Third, the irregular rhythm of AF results in irregular ventricular activation, creating the short-long-short sequences, which lead to inhomogeneous depolarization and thus to a higher myocardium susceptibility for ventricular arrhythmias. (29) FIGURE 2. Ventricular tachycardia (VT) induced by rapid ventricular response in patient with permanent atrial fibrillation (AF). Trace A: an acceleration of the ventricular rate in permanent AF. Trace B: the onset of VT with a progressive acceleration of ventricular rate. Traces C and D: established VT with very fast rate (R-R interval of 270 ms). (Adapted from reference (33)). Prognostic significance of appropriate and inappropriate implantable cardioverter defibrillator shocks Two landmark primary prevention ICD studies, MADIT-II (Multicenter Automatic Defibrillation Trial) and SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial), have found that both appropriate and inappropriate ICD shocks were significant predictors of death, increasing risk of death for appropriate shocks by 3 and 5 times, respectively, and with a 2-fold increase for inappropriate shocks. (21, 24, 34) In both studies, progressive HF was the leading cause of death among the patients who received at least one ICD shock. (24, 34) A meta-analysis by Proietti et al. (35) summarized the existing evidence on the prognostic significance of ICD shocks in patients with HFrEF. In a pooled analysis of almost 200 000 patients with ICD, a significant association was found between ICD shocks and mortality, stronger for appropriate (HR 2.95, p<0.001) than inappropriate shocks (HR 1.71, p<0.001), while the combination of both appropriate and inappropriate shocks was associated with a higher risk of death (HR 4.18, p<0.001) than either type of shock alone. How we can explain the association between ICD shocks and increased mortality? There are a number of potential mechanisms by which shocks may directly increase the risk of death, such as induced ventricular proarrhythmia, transient myocardial dysfunction, post-shock vasospasm, myocardial stunning, or post-shock pulseless electrical activity, (7, 24, 36-38) but these mechanisms can potentially be the cause of death within the first 24 hours of a shock. Another explanation arises from the fact that patients with HFrEF and more severe forms of heart disease have an increased tendency to develop ventricular arrhythmias and AF (34, 39, 40) and consequently a tendency towards a higher incidence of appropriate and inappropriate shocks, (4) which suggests the possibility that ICD shocks are indicator of a higher-risk patient. This explanation is illustrated in the study by Powel et al. (22) who analyzed the rhythms of the first shock delivery in 7 439 randomly-selected patients with an ICD or cardiac resynchronization therapy defibrillator (CRT-D) to determine their relationship with survival. Of the first shock episodes, 58.7% were appropriate shocks for sustained ventricular arrhythmias, and 41.3% were inappropriate shocks for non-ventricular arrhythmias or non-arrhythmic events. Compared with no-shock patients, those who received their first shock for VT/VF or AF/AFL had an increased risk of death (**Table 2**). In contrast, patients who received an inappropriate shock for ST/SVT or non-arrhythmic events (noise, artifact, and oversensing) had similar survival to those who did not receive a shock. Two important messages emerge from this study: 1) the risk associated with inappropriate shocks is limited to patients receiving shocks for AF/AFL, and 2) increased long-term mortality after shock is more related to the underlying arrhythmia substrate than to an adverse effect from the shock itself. Taking all this together, it is a plausible explanation that the association between ICD shocks and increased mortality is multifactorial, including more advanced heart disease, progression of the underlying arrhythmia substrate, co-morbid conditions, and, to a lesser extent, detrimental effects from the ICD shocks themselves. (22, 24, 25, 41) ### TABLE 2: Mortality risk after the first shock in comparison with no shock patients in the ALTITUDE study ( 22 ). | **Shocked rhythm** | **Hazard ratio (95% CL)** | **p value** | | --- | --- | --- | | Ventricular Monomorphic VT VF/Polymorphic VT | 1.65 (1.36-2.01) 2.10 (1.54-2.86) | <0.0001 <0.0001 | | Non-ventricular AF/AFL ST/SVT Noise/artefact/oversensing | 1.61 (1.17-2.21) 0.97 (0.68-1.37) 0.91 (0.50-1.67) | 0.003 0.86 0.76 | [†] VT = ventricular tachycardia, VF = ventricular fibrillation, AF= atrial fibrillation, AFL = atrial flutter, ST = sinus tachycardia Strategies to minimize the interference of atrial fibrillation with implantable cardioverter therapy According to recent guidelines, the optimization of ICD programming is the initial option to minimize the rate of ICD shocks induced by AF and to improve patient outcomes. (1) This recommendation is based on the results of two meta-analyses, (42, 43) which examined the overall effect of ICD therapy reduction programming on several clinical outcomes, including mortality (**Table 3**). In both meta-analyses, therapy reduction programming was associated with a 50% reduction in inappropriate shocks and with a 30% and 23% reduction in mortality, respectively, compared with the conventional programming. No increase was seen in risk of appropriate shocks or syncope in therapy reduction programming. In line with this recommendation, ICD programming should include: 1) customized bradycardia pacing mode to prevent unnecessary right ventricular pacing in patients without an indication for permanent pacing, (44, 45) 2) multi-zone tachycardia detection programming with incorporated prolonged detection time settings and high rate thresholds to avoid unnecessary shocks due to non-sustained VT or clinically stable VT, (46-48) 3) consistent use of algorithms for supraventricular versus ventricular arrhythmias discrimination even for tachycardias with rates up to 230 bpm, (17, 26, 48) 4) proper atrial sensing for activation of dual-chamber discriminators, (48) 5) systematic use of ATP before shock therapy, also for very fast VT, (46, 49, 50) and 6) specific algorithms for T-wave discrimination and right ventricular lead noise. (48) ### TABLE 3: The effect of therapy reduction versus conventional implantable cardioverter defibrillator programming on clinical outcomes ( 42 ). | **Clinical outcome** | **Therapy RP** **(n=4089)** | **Conventional P** **(n = 3598)** | **Hazard ratio** **(95% CI)** | **p value** | | --- | --- | --- | --- | --- | | All-cause mortality | 207 (5.0%) | 162 (7.3%) | 0.70 (0.59-0.84) | <0.001 | | Syncope | 105 (3.1%) | 74 (2.5%) | 1.09 (0.83-1.44) | 0.5 | | Appropriate shocks | 153 (5.2%) | 137 (5.6%) | 1.06 (0.75-1.16) | 0.5 | | Inappropriate shocks | 99 (3.4%) | 168 (6.9%) | 0.50 (0.39-0.63) | <0.001 | [†] RP = reduction programming Pharmacological therapy is the next option for the reduction of ICD shocks caused by AF, (1) but data on this issue are limited. The goal of pharmacological therapy is to slow down the heart rate so that it does not reach the detection zone for VT/VF, or to reduce AF recurrences. In a secondary analysis of the MADIT II study, (51) beta-blocker therapy (metoprolol 44%, carvedilol 42%, atenolol 13%) reduced ICD therapy for VT or VF, but failed to prevent inappropriate shocks induced by AF or SVT. On the other hand, a retrospective analysis of the MADIT-CRT (Multicenter Automatic Defibrillator Implantation with Cardiac Resynchronization Therapy) study found that the risk of inappropriate therapy due to AF was significantly reduced in patients receiving carvedilol compared with those who received metoprolol (HR: 0.50, p=0.004). (52) The superior clinical effect of carvedilol versus metoprolol was probably the result of its a larger electrophysiological and pharmacological profile. (53) In a randomized controlled trial on 412 patients with implanted ICD for secondary prevention, (54) a combination of amiodarone and beta-blocker was more effective in the prevention of inappropriate shocks (mostly for supraventricular tachyarrhythmias) than beta-blocker alone (3.3% vs. 15.4%, p=0.006), but with higher rates of adverse thyroid and pulmonary effects and of symptomatic bradycardia. On the other hand, sotalol was not more effective than beta-blockers in reducing risk of inappropriate shocks. Amiodarone can be used for rhythm control after cardioversion, but its success rate in the prevention of AF recurrences in patients with ICD or CRT-D with HFrEF and persistent AF was not higher than 34% at two-year follow-up. (55) Catheter ablation of the AF or AV junction is recommended in patients with AF-related inappropriate shocks who are unresponsive to pharmacological therapy. (1) Several studies confirmed the feasibility and usefulness of these therapeutic options. (55-58) In 2014, Kosiak et al. (56) published the results of AF catheter ablation in 73 patients with implanted ICD due to ischemic or non-ischemic cardiomyopathy. Complete pulmonary vein isolation (PVI) as the procedural endpoint was achieved in all patients, and re-ablation was necessary in 20 of them. In comparison with the period prior to ablation, there was significant reduction of appropriate and inappropriate shocks after ablation (p=0.03 and p=0.001, respectively). In the study by Miyazaki et al., (57) 14 of 106 ICD patients experienced inappropriate ICD shocks due to AF. PVI was performed in 13 patients, one patient underwent an AVJ ablation because of the absence of an inferior vena, and 4 patients were re-ablated for recurrent AF. After the last ablation procedure, no patients experienced inappropriate shock during the median follow-up of 19 months. In a pooled analysis of 664 patients with CRT-D and permanent AF, (58) patients with AVJ ablation had a significantly lower rate of inappropriate shocks induced by AF and a lower rate of appropriate shocks than those treated with drugs (both p<0.001). These data are important for the treatment of patients with CRT-D, because approximately 26% of this population have AF. (59) ## Conclusion Atrial fibrillation may interfere with ICD therapy in two ways: 1) by inducing inappropriate ICD shock when its rapid ventricular rate reaches a device’s programmed detection zone for VT/VF, and 2) by directly inducing a VT or VF episode with consequent appropriate shock therapy. Patients with ICD who receive appropriate or inappropriate shocks related to AF have a substantially higher risk of death than those who do not receive them. The association between ICD shocks and increased mortality is multifactorial, including more advanced heart disease, progression of the underlying arrhythmia substrate, co-morbid conditions, and, to a lesser extent, detrimental effects from ICD shock itself. The optimization of ICD programming is the initial option for reducing inappropriate and appropriate shocks induced by AF. Pharmacological therapy with beta-blockers for rate or amiodarone for rhythm control is the next option to this effect. The catheter ablation of AF or AV junction is recommended in patients unresponsive to pharmacological therapy depending on whether it is paroxysmal and persistent AF, or permanent AF.

    Literature

    1. Zeppenfeld K, Tfelt-Hansen J, de Riva M, Winkel BG, Behr ER, Blom NA, et al. 2022 ESC guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Developed by the Task force for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death of the European Society of Cardiology (ESC). Endorsed by the Association for European Paediatric and Congenital Cardiology (AEPC). Eur Heart J. 2022;43:3997–4126. https://doi.org/10.1093/eurheartj/ehac262
    2. Giancaterino S, Nishimura M, Birgersdotter-Green U, Hoffmayer KS, Han FT, Raissi F, et al. Clinical factors associated with baseline history of atrial fibrillp, while ation and subsequent clinical outcomes following initial implantable cardioverter-defibrillator placement. Pacing Clin Electrophysiol. 2020;43:542–50. https://doi.org/10.1111/pace.13919
    3. Maruyama M, Yasuoka R, Nagano T, Nakazawa G, Noda T, Nitta T, et al. Impact of atrial fibrillation/flutter on outcomes of patients with implantable cardioverter defibrillators: A sub-analysis of the Nippon Storm study. J Cardiol. 2021 September;78(3):244–9. https://doi.org/10.1016/j.jjcc.2021.04.003
    4. Borleffs CJW, van Rees JB, van Welesenes GH, van de Velde ET, van Erven L, Bax JJ, et al. Prognostic importance of atrial fibrillation in implantable cardioverter-defibrillator patients. J Am Coll Cardiol. 2010;55:879–85. https://doi.org/10.1016/j.jacc.2009.09.053
    5. Mustafa U, Dherange P, Reddy R, DeVillier J, Chong J, Ihsan A, et al. Atrial fibrillation is associated with higher overall mortality in patients with implantable cardioverter-defibrillator: a systematic review and meta-analysis. J Am Heart Assoc. 2018;7:e010156. https://doi.org/10.1161/JAHA.118.010156
    6. van Rees JB, Borleffs CJ, de Bie MK, Stijnen T, van Erven L, Bax JJ, et al. Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. J Am Coll Cardiol. 2011;57:556–62. https://doi.org/10.1016/j.jacc.2010.06.059
    7. Ghezzi ES, Sharman RLS, Selvanayagam JB, Psaltis PJ, Sanders P, Astley JM. Burden of mood symptoms and disorders in implantable cardioverter defibrillator patients: a systematic review and meta-analysis of 39 954 patients. Europace. 2023;25:1–15. https://doi.org/10.1093/europace/euad130
    8. Peinado R, José L, Merino JI. M. Gonzales-Vasserot. Life-threatening implantable defibrillator-induced pro-arrhythmia. Rev Esp Cardiol. 2007;60:770–1. https://doi.org/10.1157/13108282
    9. Dimitri H, John B, Young GD, Sanders P. Fatal outcome from inappropriate defibrillation. Europace. 2007;9:1059–60. https://doi.org/10.1093/europace/eum186
    10. McDonagh TA, Metra M, Adamo M, Gardner RS, Baumbsch A, Ohm M, et al. Authors/Task Force Members. 2021 ESC guidelines for the diagnosis and treatment of acute and chronic heart failure. Developed by the Task Force for diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC). With the special contribution of the Heart Failure Association (HFA) of the ESC. Eur Heart J. 2021;42:3599–726. https://doi.org/10.1093/eurheartj/ehab368
    11. Borne RT, Varosy PD, Masoudi FA. Implantable cardioverter-defibrillator shocks. Epidemiology, outcomes, and therapeutic approaches. JAMA Intern Med. 2013;173:859–65. https://doi.org/10.1001/jamainternmed.2013.428
    12. Fleeman BE, Aleong RG. Optimal strategies to reduce inappropriate implantable cardioverter-defibrillator shocks. J Innov Card Rhythm Manag. 2019;10:3623–32. https://doi.org/10.19102/icrm.2019.100403
    13. The Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmia. N Engl J Med. 1997;337:1576–83. https://doi.org/10.1056/NEJM199711273372202
    14. Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al. for the Multicenter Automatic Defibrillator Implantation Trial II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877–83. https://doi.org/10.1056/NEJMoa013474
    15. Kadish A, Dyer A, Daubert JP, Quigg R, Mark Estes NA, Anderson KP, et al. Prophylactic defibrillator implantation in patients with noniischemic dilated cardiomyopathy. N Engl J Med. 2004;350:2151–8. https://doi.org/10.1056/NEJMoa033088
    16. Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al. (SCD-HeFT) Investigators. Amiodarone or an implatable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005;352:225–37. https://doi.org/10.1056/NEJMoa043399
    17. Moss AJ, Hall J, Cannom DS, Klein H, Brown MW, Daubert JP, et al. Cardiac-resynchronization therapy for the prevention of heart failure events. N Engl J Med. 2009;361:1329–38. https://doi.org/10.1056/NEJMoa0906431
    18. Moss AJ, Schuger C, Beck C, Brown MW, Cannom DS, Daubert JP, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med. 2012;367:2275–83. https://doi.org/10.1056/NEJMoa1211107
    19. Køber L, Thune JJ, Nielsen JC, Haarbo J, Korup E, Jensen G, et al. Defibrillator Implantation in Patients with Nonischemic Systolic Heart Failure. N Engl J Med. 2016;375:1221–30. https://doi.org/10.1056/NEJMoa1608029
    20. Zweibel S, Cronin EM, Schloss EJ, Auricchio A, Kurita T, Sterns LD, et al. Estimating the incidence of atrial fibrillation in single-chamber implantable cardioverter defibrillator patients. Pacing Clin Electrophysiol. 2019;42:132–8. https://doi.org/10.1111/pace.13555
    21. Daubert JP, Zareba W, Cannom DS, McNitt S, Rosero SZ, Wag P, et al. for the MADIT II Investigators. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II. Frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol. 2008;51:1357–65. https://doi.org/10.1016/j.jacc.2007.09.073
    22. Powell BD, Saxon LA, Boehmer JP, Day JD, Gilliam FR, Heidenreich PA, et al. Survival after shock therapy in implantable cardioverter-defibrillator and cardiac resynchronization therapy-defibrillator recipients according to rhythm shocked. The ALTITUDE survival by rhythm study. J Am Coll Cardiol. 2013;62:1674–9. https://doi.org/10.1016/j.jacc.2013.04.083
    23. Klein RC, Raitt MH, Wilkoff BL, Beckman KJ, Coromilas J, Wyse G, et al. Analsyis of implantable cardioverter defibrillator therapy in the Antiarrhythmics Versus Implantable Defibrillators (AVID) trial. J Cardiovasc Electrophysiol. 2003;14:940–8. https://doi.org/10.1046/j.1540-8167.2003.01554.x
    24. Poole JE, Johnson GW, Hellkamp AS, Andersom J, Callans DJ, Raitt MH, et al. Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med. 2008;359:1009–17. https://doi.org/10.1056/NEJMoa071098
    25. Sood N, Ruwald AC, Solomon S, Daubert JP, McNitt S, Polonsky B, et al. Association between myocardial substrate, implantable cardioverter defibrillator shocks and mortality in MADIT-CRT. Eur Heart J. 2014;35:106–15. https://doi.org/10.1093/eurheartj/eht451
    26. Auricchio A, Schloss EJ, Kurita T, Meijer A, Gerritse B, Zweibel S, et al. On behalf of the PainFree SST Investigators. Low inappropriate shock rates in patients with single- and dual/triple-chamber implantable cardioverter-defibrillators using a novel suite of detection algorithms: PainFree SST trial primary results. Heart Rhythm. 2015 May;12(5):926–36. https://doi.org/10.1016/j.hrthm.2015.01.017
    27. Ruiz-Granell R, Dovellini EV, Dompnier A, Khatighi K, Garcia-Campo E, Olivier A, et al. Algorithm-based reduction of inapropriate defibrillator shock: results of the Inappropriate Shock with PARAD+ Rhythm DiScrimination-Implantable Cardioverter Defibrillator Study. Heart Rhythm. 2019 September;16(9):1429–35. https://doi.org/10.1016/j.hrthm.2019.03.016
    28. Marchlinski FE, Callans DJ, Gottlieb C, Schwartzman D. MARK Preminger M. Benefits and lessons learned from stored electrogram information in implantable defibrillators. J Cardiovasc Electrophysiol. 1995;6:832–51. https://doi.org/10.1111/j.1540-8167.1995.tb00359.x
    29. Grönefeld GC, Mauss O, Li YG, Klingenheben T, Hohnloser SH. Association between atrial fibrillation and appropriate implantable cardioverter defibrillator therapy: results from a prospective study. J Cardiovasc Electrophysiol. 2000;11:1208–14. https://doi.org/10.1046/j.1540-8167.2000.01208.x
    30. Stein KM, Euler DE, Mehra R, Seidl K, Slotwiner DJ, Mittal S, et al. Do atrial tachyarrhythmias beget ventricular tachyarrhythmias in defibrillator recipients. J Am Coll Cardiol. 2002;40:335–40. https://doi.org/10.1016/S0735-1097(02)01957-5
    31. Smit MD, Van Dessel PFHM, Rienstra M, Nieuwland W, Wiesfeld ACP, Tan ES, et al. Atrial fibrillation predicts appropriate shocks in primary prevention implantable cardioverter-defibrillator patients. Europace. 2006;8:566–72. https://doi.org/10.1093/europace/eul081
    32. Sakhuja R, Shah AJ, Keebler M, Thakur RK. Atrial fibrillation in patients with implantable defibrillators. Cardiol Clin. 2009;27:151–61. https://doi.org/10.1016/j.ccl.2008.09.014
    33. Petrac D, Radićc B, Radeljić V, Hamel D, Filipović J. Impact of atrioventricular node ablation and pacing therapy on clinical course in patients with permanent atrial fibrillation and unstable ventricular tachycardia induced by rapid ventricular response: follow-up study. Croat Med J. 2005 Dec;46(6):929-35. https://pubmed.ncbi.nlm.nih.gov/16342346/
    34. Moss AJ, Greenberg H, Case RB, Zareba W, Hall J, Brown MW, et al. Long-term clinical course of patients after termination of ventricular tachyarrhythmia by an Implanted Defibrillator. Circulation. 2004;110:3760–5. https://doi.org/10.1161/01.CIR.0000150390.04704.B7
    35. Proietti R, Labos C, Davis M, Thanassoulis G, Santaganeli P, Russo V, et al. A systematic review and meta-analysis of the association between implantable cardioverter-defibrillator shocks and long-term mortality. Can J Cardiol. 2015;31:270–7. https://doi.org/10.1016/j.cjca.2014.11.023
    36. Mitchell LB, Pineda EA, Titus JL, Bartosch PM, Beneditt DG. Sudden death in patients with implantable cardioverter defibrillators. The importance of post-shock electromechanical dissociation. J Am Coll Cardiol. 2002;39:1323–8. https://doi.org/10.1016/S0735-1097(02)01784-9
    37. Clementy N, Bodin A, Bisson A, Teixeira-Gomez AP, Roge S, Angouvlant D, et al. The defibrillation conundrum: new insights into the mechanisms of shock-related myocardial injury sustained from a life-saving therapy. Int J Mol Sci. 2021;22:1–17. https://doi.org/10.3390/ijms22095003
    38. Tereshchenko LG, Faddis MN, Fetics BJ, Zelik KE, Efimov IR, Berger RD. Transient local injury current in right ventricular electrogram after implantable cardioverter-defibrillator shock predicts heart failure progression. J Am Coll Cardiol. 2009;54:822–8. https://doi.org/10.1016/j.jacc.2009.06.004
    39. Tan NY, Roger VL, Killian JM, Cha YM, Noseworthy PA, Dunlay SM. Ventricular arrhythmias among patients with advanced heart failure: a population‐based study. J Am Heart Assoc. 2022;11:e023377. https://doi.org/10.1161/JAHA.121.023377
    40. Santhanakrishnan R, Wang NA, Larsen MG, Mgnani JW, McManus DD, Lubitz SA, et al. Atrial fibrillation begets heart failure and vice versa: Temporal associations and differences in preserved versus reduced ejection fraction. Circulation. 2016;133:484–92. https://doi.org/10.1161/CIRCULATIONAHA.115.018614
    41. Aktaş MK, Younis A, Zareba W, Kutyifa V, Klein H, Daubert JP, et al. Survival after implantable cardioverter defibrillator shocks. J Am Coll Cardiol. 2021;77:2453–62. https://doi.org/10.1016/j.jacc.2021.03.329
    42. Tan VH, Wilton SB, Kuriachan V, Sumner GL, Exner DV. Impact of programming strategies aimed at reducing nonessential implantable cardioverter defibrillator therapies on mortality. A systematic review and meta-analysis. Circ Arrhythm Electrophysiol. 2014;7:164–70. https://doi.org/10.1161/CIRCEP.113.001217
    43. Scott PA, Silberbauer J, McDonagh TA, Murgatroyd FD. Impact of prolonged implantable cardioverter-defibrillator arrhythmia detection times on outcomes: a meta-analysis. Heart Rhythm. 2014;11:828–35. https://doi.org/10.1016/j.hrthm.2014.02.009
    44. Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA. 2002;288:3115–23. https://doi.org/10.1001/jama.288.24.3115
    45. Barsheshet A, Moss AJ, McNitt S, Jons C, Glikson M, Klein HU, et al. Long-term implications of cumulative right ventricular pacing among patients with an implantable cardioverter-defibrillator. Heart Rhythm. 2011;8:212–8. https://doi.org/10.1016/j.hrthm.2010.10.035
    46. Gasparini M, Proclemer A, Klersy C, Kloppe A, Ferrer JBM, Hersi A, et al. Effect of long-detection interval vs standard-detection interval for implantable cardioverter defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA. 2013;309:1903–11. https://doi.org/10.1001/jama.2013.4598
    47. Saeed M, Hanna I, Robotis D, Styperek R, Polosajian L, Khan A, et al. Programming implantable cardioverter-defibrillators in patients with primary prevention indication to prolong time to first shock: results from the PROVIDE study. J Cardiovasc Electrophysiol. 2014;25:52–9. https://doi.org/10.1111/jce.12273
    48. Wilkoff BL, Fauchier L, Stiles MK, Morillo CA, Al-Khatib SM, Almendral J, et al. 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Europace. 2016;18:159–83. https://doi.org/10.1093/europace/euv411
    49. Wilkoff BL, Ousdigian KT, Sterns LD, Wang ZJ, Wilson RD, Morgan JM, et al. A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicentre EMPIRIC trial. J Am Coll Cardiol. 2006;48:330–9. https://doi.org/10.1016/j.jacc.2006.03.037
    50. Wathen MS, DeGroot PJ, Sweeney MO, Stark AJ, Otterness MF, Adkisson WO, et al. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. Circulation. 2004;110:2591–6. https://doi.org/10.1161/01.CIR.0000145610.64014.E4
    51. Brodine WN, Tung RT, Lee JK, Hockstad ES, Moss AJ, Zareba W, et al. Effects of Beta-Blockers on Implantable Cardioverter Defibrillator Therapy and Survival in the Patients with Ischemic Cardiomyopathy (from the Multicenter Automatic Defibrillator Implantation Trial-II). Am J Cardiol. 2005;96:691–5. https://doi.org/10.1016/j.amjcard.2005.04.046
    52. Ruwald MH, Abu-Zeitone A, Jons C, Ruwald AC, McNitt S, Kutyifa V, et al. Impact of carvedilol and metoprolol on inappropriate implantable cardioverter-defibrillator therapy: the MADIT-CRT trial (Multicenter Automatic Defibrillator Implantation with Cardiac Resynchronization Therapy). J Am Coll Cardiol. 2013;62:1343–50. https://doi.org/10.1016/j.jacc.2013.03.087
    53. Metra M, Giubbini R, Nodari S, Boldi E, Moden MG, Dei CASL. Differential Effects of beta-blockers in patients with heart failure. A prospective, randomized, double-blind comparison of the long-term effects of metoprolol versus carvedilol. Circulation. 2000;102:546–51. https://doi.org/10.1161/01.CIR.102.5.546
    54. Connolly SJ, Dorian P, Roberts RS, Gent M, Bailin S, Fain ES, et al. Comparison of beta-blockers, amiodarone plus beta-blockers, or sotalol for prevention of shocks from implantable cardioverter defibrillators: the OPTIC Study: a randomized trial. JAMA. 2006;295:165–71. https://doi.org/10.1001/jama.295.2.165
    55. Di Biase L, Mohanty P, Mohanty S, Santangeli P, Trivedi C, Lakkireddy D, et al. Ablation versus amiodarone for treatment of persistent atrial fibrillation in patients with congestive heart failure and an implanted device. Results from the AATAC multicenter randomized trial. Circulation. 2016;133:1637–44. https://doi.org/10.1161/CIRCULATIONAHA.115.019406
    56. Kosiuk J, Nedios S, Darma A, Rolf S, Richter S, Arya A, et al. Impact of single atrial fibrillation catheter ablation on implantable cardioverter defibrillator therapies in patients with ischaemic and non-ischaemic cardiomyopathies. Europace. 2014;16:1322–6. https://doi.org/10.1093/europace/euu018
    57. Miyazaki S, Taniguchi H, Kusa S, Komatsu Y, Ichihara N, Takagi T, et al. Catheter ablation of atrial tachyarrhythmias causing inappropriate implantable cardioverter-defibrillator shocks. Europace. 2015;17:289–94. https://doi.org/10.1093/europace/euu185
    58. Gasparini M, Kloppe A, Lunati M, Anselme F, Landolina M, Martinez-Ferrer JB, et al. Atrioventricular junction ablation in patients with atrial fibrillation treated with cardiac resynchronization therapy: positive impact on ventricular arrhythmias, implantable cardioverter-defibrillator therapies and hospitalizations: atrioventricular junction ablation in CRT patients with AF. Eur J Heart Fail. 2018;20:1472–81. https://doi.org/10.1002/ejhf.1117
    59. Dickstein K, Normand C, Auricchio A, Bogale N, Cleland JG, Gitt A, et al. CRT Survey II: a European Society of Cardiology survey of cardiac resynchronisation therapy in 11088 patients - who is doing what to whom and how? Eur J Heart Fail. 2018;20:1039–51. https://doi.org/10.1002/ejhf.1142
    Cardiologia Croatica
    Back to search

    Interference of Atrial Fibrillation with Implantable Cardioverter Defibrillator Therapy in Patients with Heart Failure and Reduced Ejection Fraction

    Review Article
    Issue3-4
    Published
    Pages177-185
    PDF via DOIhttps://doi.org/10.15836/ccar2024.177
    interferencija
    fibrilacija atrija
    implantabilni kardioverterski defibrilator
    interference
    atrial fibrillation
    implantable cardioverter defibrillator

    Authors

    Dubravko Petrač*Croatia Poliklinika, Zagreb, Hrvatska
    Vjekoslav RadeljićKlinički bolnički centar Sestre milosrdnice, Zagreb, Hrvatska
    Diana Delić-BrkljačićKlinički bolnički centar Sestre milosrdnice, Zagreb, Hrvatska

    *Correspondence email: d.petrac@inet.hr

    Abstract

    Implantable cardioverter defibrillators (ICD) are an important therapeutic option in reducing mortality due to ventricular arrhythmias in patients with heart failure with reduced ejection fraction. Atrial fibrillation (AF) is often present in these patients and may interfere with ICD therapy by inducing inappropriate and appropriate shocks. This issue is important because both types of shock increase mortality in patients with ICD. The strategies to minimize the rate of ICD shocks induced by AF include the optimization of ICD programming, pharmacological therapy for rate or rhythm control, and ablation of AF or atrioventricular junction ablation. In this review, we describe the interference of AF with ICD therapy, examine the impact of this interference on survival, and discuss the strategies for its reduction.

    Full Text

    Introduction

    Implantable cardioverter defibrillators (ICD) are a well-established and effective therapy for patients with heart failure with reduced ejection fraction (HFrEF) who survived sustained ventricular tachycardia (VT) or ventricular fibrillation (VF), or those who are exposed to an increased risk of these arrhythmias. (1) With regard to poor left ventricular systolic function, AF is often present in patients with ICD, with a prevalence from 17% to 36% at time of implantation in clinical practice. (2, 3) Except for the negative effect on survival, (4, 5) AF may interfere with ICD therapy and limits its clinical usefulness by inducing inappropriate and appropriate shocks. This issue is important because ICD shocks are painful, cause psychological disturbances, and may induce ventricular pro-arrhythmia and even death. (6–9) The aim of this review was to explore the possible interference mechanisms of AF with ICD therapy, examine the impact of this interference on survival, and discuss the treatment options available for its reduction.

    Mechanisms of atrial fibrillation interference with implantable cardioverter-defibrillator therapy

    ICDs are an important therapeutic option in reducing mortality due to ventricular arrhythmias in patients with HFrEF. (10) The primary goal of ICD therapy is to detect and stop the life-threatening ventricular arrhythmia by delivering electrical shock and thus prevent a sudden cardiac death. In that context, ICD shocks delivered for VT/VF are considered appropriate because they save lives. (11) In contrast, ICD shocks delivered for the non-ventricular arrhythmias, like AF, atrial flutter (AFL), supraventricular tachycardia (SVT), atrial tachycardia (AT), sinus tachycardia (ST), or non-arrhythmic events, like electromagnetic interference, myopotentials, and device malfunction or oversensing, are unnecessary and are considered inappropriate. (12)

    AF may interfere with ICD therapy in two ways. The first way is more common, and occurs when the fast ventricular rate of AF with relatively regular R-R intervals reaches a device’s programmed detection zone for VT/VF, is misclassified as ventricular arrhythmia, and induces the delivery of inappropriate shock (Figure 1). According to data from randomized clinical trials, a history of AF was found in 9% to 24% of patients with ICD (Table 1), more often in patients with non-ischemic than in those with ischemic cardiomyopathy. (13–19) On the other hand, the incidence of new-onset AF lasting at least 24 hours per day was 6.3% and 7.1% in single chamber and doubled chamber patients with ICD, respectively. (20) Therefore, it is not surprising that AF/AFL was the most common reason for an inappropriate ICD shock, followed by ST/SVT and non-arrhythmic events. (21, 22) In earlier randomized clinical trials, (15, 21, 23, 24) the rate of inappropriate ICD shocks ranged from 15% to 21% over 20 to 45.5 months of follow-up (Table 1). Later, thanks to device programming adjustment and enhanced algorithm discrimination of arrhythmias, the rate of inappropriate shocks decreased, (18, 19, 25) but even with these improvements, they occur in 1.9% to 3.7% of patients with ICD over a period of two years. (26, 27)

    FIGURE 1. Inappropriate shock induced by atrial fibrillation in a patient with a double-chamber implantable cardioverter defibrillator. AS = ATRIAL SENSING, VS = VENTRICULAR SENSING, CD = CARDIOVERSION.

    TABLE 1: Summary of atrial fibrillation history and implantable cardioverter-defibrillator shocks in randomized controlled trials.

    AVID, (13, 23)
    No of ICD patients
    507
    Heart disease (%)
    ICM 81 DICM 19
    History of AF (%)
    21
    EF (median)
    32%
    FU (m) (mean)
    18
    ICD shocks (%) Approp Inapprop
    39 20
    MADIT II (14, 21)
    No of ICD patients
    719
    Heart disease (%)
    ICM 100
    History of AF (%)
    9
    EF (median)
    23%
    FU (m) (mean)
    20
    ICD shocks (%) Approp Inapprop
    18 15
    DEFINITE (15)
    No of ICD patients
    229
    Heart disease (%)
    DICM 100
    History of AF (%)
    23
    EF (median)
    29%
    FU (m) (mean)
    29
    ICD shocks (%) Approp Inapprop
    18 21
    SCD-HeFT (16, 24)
    No of ICD patients
    811
    Heart disease (%)
    ICM 53 DICM 47
    History of AF (%)
    17
    EF (median)
    25%
    FU (m) (mean)
    45
    ICD shocks (%) Approp Inapprop
    22 17
    MADIT-CRT (17, 25)
    No of ICD patients
    1790
    Heart disease (%)
    ICM 55 DICM 45
    History of AF (%)
    11
    EF (median)
    28%
    FU (m) (mean)
    39
    ICD shocks (%) Approp Inapprop
    13 7
    MADIT-RIT (18)
    No of ICD patients
    1500
    Heart disease (%)
    ICM 53 DICM 47
    History of AF (%)
    13
    EF (median)
    26%
    FU (m) (mean)
    17
    ICD shocks (%) Approp Inapprop
    13 4
    DANISH (19)
    No of ICD patients
    556
    Heart disease (%)
    DICM 100
    History of AF (%)
    24
    EF (median)
    25%
    FU (m) (mean)
    67
    ICD shocks (%) Approp Inapprop
    12 6

    RCT = randomized controlled trial, No = number, AF = atrial fibrillation, EF = ejection fraction, FU = follow-up, m = months, ICD = implantable cardioverter-defibrillator, approp = appropriate, inapprop = inappropriate, ISM = ischemic cardiomyopathy, NICM = non-ischemic cardiomyopathy.

    The second way in which AF may interfere with ICD therapy is direct action of AF in causing an episode of VT/VF, with consequent appropriate shock therapy. An earlier analysis of stored ICD electrograms revealed that approximately 10% of VT episodes were preceded by atrial fibrillation typically associated with a rapid ventricular response. (28) Subsequently, Gronefeld et al. (29) reported that patients in AF experienced appropriate ICD therapy for recurrent ventricular arrhythmias more frequently than those in sinus rhythm (63% vs. 38%, p=0.01) because of a higher incidence of short-long-short cycles preceding ventricular arrhythmias in AF (50% vs. 16%, p=0.002). In the study by Stein et al. (30) which included 537 patients with dual-chamber ICD, 233 (8.6%) of 2602 VT/VF episodes were preceded by a paroxysm of AT or AF. The median duration of these arrhythmias preceding VT/VF was approximately one hour. Ventricular therapies were delivered to treat 205 of these VT/VF episodes, while the remaining 28 episodes terminated spontaneously before an ICD therapy could be delivered. On the other hand, Borleffs et al. reported that patients with permanent AF had higher risk of ventricular arrhythmias triggering ICD discharge compared with those without AF and those with paroxysmal or persistent AF (49% vs. 29%, 26%, and 26%, respectively; p<0.001). (4)

    There are several mechanisms that can explain the association between AF and appropriate ICD shocks. First, it is possible that AF and ventricular arrhythmias share risk factors such as ischemia, renal failure, increased sympathetic tone, or increased left ventricular filling pressure, (31, 32) which facilitate the initiation of both arrhythmias. Second, in structural heart disease with reduced EF, the rapid ventricular response caused by AF may directly affect ventricular refractoriness and therefore induce ventricular arrhythmia (Figure 2). (30, 33) Third, the irregular rhythm of AF results in irregular ventricular activation, creating the short-long-short sequences, which lead to inhomogeneous depolarization and thus to a higher myocardium susceptibility for ventricular arrhythmias. (29)

    FIGURE 2. Ventricular tachycardia (VT) induced by rapid ventricular response in patient with permanent atrial fibrillation (AF). Trace A: an acceleration of the ventricular rate in permanent AF. Trace B: the onset of VT with a progressive acceleration of ventricular rate. Traces C and D: established VT with very fast rate (R-R interval of 270 ms). (Adapted from reference (33)).

    Prognostic significance of appropriate and inappropriate implantable cardioverter defibrillator shocks

    Two landmark primary prevention ICD studies, MADIT-II (Multicenter Automatic Defibrillation Trial) and SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial), have found that both appropriate and inappropriate ICD shocks were significant predictors of death, increasing risk of death for appropriate shocks by 3 and 5 times, respectively, and with a 2-fold increase for inappropriate shocks. (21, 24, 34) In both studies, progressive HF was the leading cause of death among the patients who received at least one ICD shock. (24, 34) A meta-analysis by Proietti et al. (35) summarized the existing evidence on the prognostic significance of ICD shocks in patients with HFrEF. In a pooled analysis of almost 200 000 patients with ICD, a significant association was found between ICD shocks and mortality, stronger for appropriate (HR 2.95, p<0.001) than inappropriate shocks (HR 1.71, p<0.001), while the combination of both appropriate and inappropriate shocks was associated with a higher risk of death (HR 4.18, p<0.001) than either type of shock alone.

    How we can explain the association between ICD shocks and increased mortality? There are a number of potential mechanisms by which shocks may directly increase the risk of death, such as induced ventricular proarrhythmia, transient myocardial dysfunction, post-shock vasospasm, myocardial stunning, or post-shock pulseless electrical activity, (7, 24, 36–38) but these mechanisms can potentially be the cause of death within the first 24 hours of a shock. Another explanation arises from the fact that patients with HFrEF and more severe forms of heart disease have an increased tendency to develop ventricular arrhythmias and AF (34, 39, 40) and consequently a tendency towards a higher incidence of appropriate and inappropriate shocks, (4) which suggests the possibility that ICD shocks are indicator of a higher-risk patient. This explanation is illustrated in the study by Powel et al. (22) who analyzed the rhythms of the first shock delivery in 7 439 randomly-selected patients with an ICD or cardiac resynchronization therapy defibrillator (CRT-D) to determine their relationship with survival. Of the first shock episodes, 58.7% were appropriate shocks for sustained ventricular arrhythmias, and 41.3% were inappropriate shocks for non-ventricular arrhythmias or non-arrhythmic events. Compared with no-shock patients, those who received their first shock for VT/VF or AF/AFL had an increased risk of death (Table 2). In contrast, patients who received an inappropriate shock for ST/SVT or non-arrhythmic events (noise, artifact, and oversensing) had similar survival to those who did not receive a shock. Two important messages emerge from this study: 1) the risk associated with inappropriate shocks is limited to patients receiving shocks for AF/AFL, and 2) increased long-term mortality after shock is more related to the underlying arrhythmia substrate than to an adverse effect from the shock itself. Taking all this together, it is a plausible explanation that the association between ICD shocks and increased mortality is multifactorial, including more advanced heart disease, progression of the underlying arrhythmia substrate, co-morbid conditions, and, to a lesser extent, detrimental effects from the ICD shocks themselves. (22, 24, 25, 41)

    TABLE 2: Mortality risk after the first shock in comparison with no shock patients in the ALTITUDE study (22)

    Ventricular Monomorphic VT VF/Polymorphic VT
    Hazard ratio (95% CL)
    1.65 (1.36-2.01) 2.10 (1.54-2.86)
    p value
    <0.0001 <0.0001
    Non-ventricular AF/AFL ST/SVT Noise/artefact/oversensing
    Hazard ratio (95% CL)
    1.61 (1.17-2.21) 0.97 (0.68-1.37) 0.91 (0.50-1.67)
    p value
    0.003 0.86 0.76

    VT = ventricular tachycardia, VF = ventricular fibrillation, AF= atrial fibrillation, AFL = atrial flutter, ST = sinus tachycardia

    Strategies to minimize the interference of atrial fibrillation with implantable cardioverter therapy

    According to recent guidelines, the optimization of ICD programming is the initial option to minimize the rate of ICD shocks induced by AF and to improve patient outcomes. (1) This recommendation is based on the results of two meta-analyses, (42, 43) which examined the overall effect of ICD therapy reduction programming on several clinical outcomes, including mortality (Table 3). In both meta-analyses, therapy reduction programming was associated with a 50% reduction in inappropriate shocks and with a 30% and 23% reduction in mortality, respectively, compared with the conventional programming. No increase was seen in risk of appropriate shocks or syncope in therapy reduction programming. In line with this recommendation, ICD programming should include: 1) customized bradycardia pacing mode to prevent unnecessary right ventricular pacing in patients without an indication for permanent pacing, (44, 45) 2) multi-zone tachycardia detection programming with incorporated prolonged detection time settings and high rate thresholds to avoid unnecessary shocks due to non-sustained VT or clinically stable VT, (46–48) 3) consistent use of algorithms for supraventricular versus ventricular arrhythmias discrimination even for tachycardias with rates up to 230 bpm, (17, 26, 48) 4) proper atrial sensing for activation of dual-chamber discriminators, (48) 5) systematic use of ATP before shock therapy, also for very fast VT, (46, 49, 50) and 6) specific algorithms for T-wave discrimination and right ventricular lead noise. (48)

    TABLE 3: The effect of therapy reduction versus conventional implantable cardioverter defibrillator programming on clinical outcomes (42)

    All-cause mortality
    Therapy RP (n=4089)
    207 (5.0%)
    Conventional P (n = 3598)
    162 (7.3%)
    Hazard ratio (95% CI)
    0.70 (0.59-0.84)
    p value
    <0.001
    Syncope
    Therapy RP (n=4089)
    105 (3.1%)
    Conventional P (n = 3598)
    74 (2.5%)
    Hazard ratio (95% CI)
    1.09 (0.83-1.44)
    p value
    0.5
    Appropriate shocks
    Therapy RP (n=4089)
    153 (5.2%)
    Conventional P (n = 3598)
    137 (5.6%)
    Hazard ratio (95% CI)
    1.06 (0.75-1.16)
    p value
    0.5
    Inappropriate shocks
    Therapy RP (n=4089)
    99 (3.4%)
    Conventional P (n = 3598)
    168 (6.9%)
    Hazard ratio (95% CI)
    0.50 (0.39-0.63)
    p value
    <0.001

    RP = reduction programming

    Pharmacological therapy is the next option for the reduction of ICD shocks caused by AF, (1) but data on this issue are limited. The goal of pharmacological therapy is to slow down the heart rate so that it does not reach the detection zone for VT/VF, or to reduce AF recurrences. In a secondary analysis of the MADIT II study, (51) beta-blocker therapy (metoprolol 44%, carvedilol 42%, atenolol 13%) reduced ICD therapy for VT or VF, but failed to prevent inappropriate shocks induced by AF or SVT. On the other hand, a retrospective analysis of the MADIT-CRT (Multicenter Automatic Defibrillator Implantation with Cardiac Resynchronization Therapy) study found that the risk of inappropriate therapy due to AF was significantly reduced in patients receiving carvedilol compared with those who received metoprolol (HR: 0.50, p=0.004). (52) The superior clinical effect of carvedilol versus metoprolol was probably the result of its a larger electrophysiological and pharmacological profile. (53) In a randomized controlled trial on 412 patients with implanted ICD for secondary prevention, (54) a combination of amiodarone and beta-blocker was more effective in the prevention of inappropriate shocks (mostly for supraventricular tachyarrhythmias) than beta-blocker alone (3.3% vs. 15.4%, p=0.006), but with higher rates of adverse thyroid and pulmonary effects and of symptomatic bradycardia. On the other hand, sotalol was not more effective than beta-blockers in reducing risk of inappropriate shocks. Amiodarone can be used for rhythm control after cardioversion, but its success rate in the prevention of AF recurrences in patients with ICD or CRT-D with HFrEF and persistent AF was not higher than 34% at two-year follow-up. (55)

    Catheter ablation of the AF or AV junction is recommended in patients with AF-related inappropriate shocks who are unresponsive to pharmacological therapy. (1) Several studies confirmed the feasibility and usefulness of these therapeutic options. (55–58) In 2014, Kosiak et al. (56) published the results of AF catheter ablation in 73 patients with implanted ICD due to ischemic or non-ischemic cardiomyopathy. Complete pulmonary vein isolation (PVI) as the procedural endpoint was achieved in all patients, and re-ablation was necessary in 20 of them. In comparison with the period prior to ablation, there was significant reduction of appropriate and inappropriate shocks after ablation (p=0.03 and p=0.001, respectively). In the study by Miyazaki et al., (57) 14 of 106 ICD patients experienced inappropriate ICD shocks due to AF. PVI was performed in 13 patients, one patient underwent an AVJ ablation because of the absence of an inferior vena, and 4 patients were re-ablated for recurrent AF. After the last ablation procedure, no patients experienced inappropriate shock during the median follow-up of 19 months. In a pooled analysis of 664 patients with CRT-D and permanent AF, (58) patients with AVJ ablation had a significantly lower rate of inappropriate shocks induced by AF and a lower rate of appropriate shocks than those treated with drugs (both p<0.001). These data are important for the treatment of patients with CRT-D, because approximately 26% of this population have AF. (59)

    Conclusion

    Atrial fibrillation may interfere with ICD therapy in two ways: 1) by inducing inappropriate ICD shock when its rapid ventricular rate reaches a device’s programmed detection zone for VT/VF, and 2) by directly inducing a VT or VF episode with consequent appropriate shock therapy. Patients with ICD who receive appropriate or inappropriate shocks related to AF have a substantially higher risk of death than those who do not receive them. The association between ICD shocks and increased mortality is multifactorial, including more advanced heart disease, progression of the underlying arrhythmia substrate, co-morbid conditions, and, to a lesser extent, detrimental effects from ICD shock itself.

    The optimization of ICD programming is the initial option for reducing inappropriate and appropriate shocks induced by AF. Pharmacological therapy with beta-blockers for rate or amiodarone for rhythm control is the next option to this effect. The catheter ablation of AF or AV junction is recommended in patients unresponsive to pharmacological therapy depending on whether it is paroxysmal and persistent AF, or permanent AF.

    Literature

    1. 1.
      Zeppenfeld K, Tfelt-Hansen J, de Riva M, Winkel BG, Behr ER, Blom NA, et al. 2022 ESC guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Developed by the Task force for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death of the European Society of Cardiology (ESC). Endorsed by the Association for European Paediatric and Congenital Cardiology (AEPC). Eur Heart J. 2022;43:3997–4126.DOI
    2. 2.
      Giancaterino S, Nishimura M, Birgersdotter-Green U, Hoffmayer KS, Han FT, Raissi F, et al. Clinical factors associated with baseline history of atrial fibrillp, while ation and subsequent clinical outcomes following initial implantable cardioverter-defibrillator placement. Pacing Clin Electrophysiol. 2020;43:542–50.DOI
    3. 3.
      Maruyama M, Yasuoka R, Nagano T, Nakazawa G, Noda T, Nitta T, et al. Impact of atrial fibrillation/flutter on outcomes of patients with implantable cardioverter defibrillators: A sub-analysis of the Nippon Storm study. J Cardiol. 2021 September;78(3):244–9.DOI
    4. 4.
      Borleffs CJW, van Rees JB, van Welesenes GH, van de Velde ET, van Erven L, Bax JJ, et al. Prognostic importance of atrial fibrillation in implantable cardioverter-defibrillator patients. J Am Coll Cardiol. 2010;55:879–85.DOI
    5. 5.
      Mustafa U, Dherange P, Reddy R, DeVillier J, Chong J, Ihsan A, et al. Atrial fibrillation is associated with higher overall mortality in patients with implantable cardioverter-defibrillator: a systematic review and meta-analysis. J Am Heart Assoc. 2018;7:e010156.DOI
    6. 6.
      van Rees JB, Borleffs CJ, de Bie MK, Stijnen T, van Erven L, Bax JJ, et al. Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. J Am Coll Cardiol. 2011;57:556–62.DOI
    7. 7.
      Ghezzi ES, Sharman RLS, Selvanayagam JB, Psaltis PJ, Sanders P, Astley JM. Burden of mood symptoms and disorders in implantable cardioverter defibrillator patients: a systematic review and meta-analysis of 39 954 patients. Europace. 2023;25:1–15.DOI
    8. 8.
      Peinado R, José L, Merino JI. M. Gonzales-Vasserot. Life-threatening implantable defibrillator-induced pro-arrhythmia. Rev Esp Cardiol. 2007;60:770–1.DOI
    9. 9.
      Dimitri H, John B, Young GD, Sanders P. Fatal outcome from inappropriate defibrillation. Europace. 2007;9:1059–60.DOI
    10. 10.
      McDonagh TA, Metra M, Adamo M, Gardner RS, Baumbsch A, Ohm M, et al. Authors/Task Force Members. 2021 ESC guidelines for the diagnosis and treatment of acute and chronic heart failure. Developed by the Task Force for diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC). With the special contribution of the Heart Failure Association (HFA) of the ESC. Eur Heart J. 2021;42:3599–726.DOI
    11. 11.
      Borne RT, Varosy PD, Masoudi FA. Implantable cardioverter-defibrillator shocks. Epidemiology, outcomes, and therapeutic approaches. JAMA Intern Med. 2013;173:859–65.DOI
    12. 12.
      Fleeman BE, Aleong RG. Optimal strategies to reduce inappropriate implantable cardioverter-defibrillator shocks. J Innov Card Rhythm Manag. 2019;10:3623–32.DOI
    13. 13.
      The Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmia. N Engl J Med. 1997;337:1576–83.DOI
    14. 14.
      Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al. for the Multicenter Automatic Defibrillator Implantation Trial II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877–83.DOI
    15. 15.
      Kadish A, Dyer A, Daubert JP, Quigg R, Mark Estes NA, Anderson KP, et al. Prophylactic defibrillator implantation in patients with noniischemic dilated cardiomyopathy. N Engl J Med. 2004;350:2151–8.DOI
    16. 16.
      Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al. (SCD-HeFT) Investigators. Amiodarone or an implatable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005;352:225–37.DOI
    17. 17.
      Moss AJ, Hall J, Cannom DS, Klein H, Brown MW, Daubert JP, et al. Cardiac-resynchronization therapy for the prevention of heart failure events. N Engl J Med. 2009;361:1329–38.DOI
    18. 18.
      Moss AJ, Schuger C, Beck C, Brown MW, Cannom DS, Daubert JP, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med. 2012;367:2275–83.DOI
    19. 19.
      Køber L, Thune JJ, Nielsen JC, Haarbo J, Korup E, Jensen G, et al. Defibrillator Implantation in Patients with Nonischemic Systolic Heart Failure. N Engl J Med. 2016;375:1221–30.DOI
    20. 20.
      Zweibel S, Cronin EM, Schloss EJ, Auricchio A, Kurita T, Sterns LD, et al. Estimating the incidence of atrial fibrillation in single-chamber implantable cardioverter defibrillator patients. Pacing Clin Electrophysiol. 2019;42:132–8.DOI
    21. 21.
      Daubert JP, Zareba W, Cannom DS, McNitt S, Rosero SZ, Wag P, et al. for the MADIT II Investigators. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II. Frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol. 2008;51:1357–65.DOI
    22. 22.
      Powell BD, Saxon LA, Boehmer JP, Day JD, Gilliam FR, Heidenreich PA, et al. Survival after shock therapy in implantable cardioverter-defibrillator and cardiac resynchronization therapy-defibrillator recipients according to rhythm shocked. The ALTITUDE survival by rhythm study. J Am Coll Cardiol. 2013;62:1674–9.DOI
    23. 23.
      Klein RC, Raitt MH, Wilkoff BL, Beckman KJ, Coromilas J, Wyse G, et al. Analsyis of implantable cardioverter defibrillator therapy in the Antiarrhythmics Versus Implantable Defibrillators (AVID) trial. J Cardiovasc Electrophysiol. 2003;14:940–8.DOI
    24. 24.
      Poole JE, Johnson GW, Hellkamp AS, Andersom J, Callans DJ, Raitt MH, et al. Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med. 2008;359:1009–17.DOI
    25. 25.
      Sood N, Ruwald AC, Solomon S, Daubert JP, McNitt S, Polonsky B, et al. Association between myocardial substrate, implantable cardioverter defibrillator shocks and mortality in MADIT-CRT. Eur Heart J. 2014;35:106–15.DOI
    26. 26.
      Auricchio A, Schloss EJ, Kurita T, Meijer A, Gerritse B, Zweibel S, et al. On behalf of the PainFree SST Investigators. Low inappropriate shock rates in patients with single- and dual/triple-chamber implantable cardioverter-defibrillators using a novel suite of detection algorithms: PainFree SST trial primary results. Heart Rhythm. 2015 May;12(5):926–36.DOI
    27. 27.
      Ruiz-Granell R, Dovellini EV, Dompnier A, Khatighi K, Garcia-Campo E, Olivier A, et al. Algorithm-based reduction of inapropriate defibrillator shock: results of the Inappropriate Shock with PARAD+ Rhythm DiScrimination-Implantable Cardioverter Defibrillator Study. Heart Rhythm. 2019 September;16(9):1429–35.DOI
    28. 28.
      Marchlinski FE, Callans DJ, Gottlieb C, Schwartzman D. MARK Preminger M. Benefits and lessons learned from stored electrogram information in implantable defibrillators. J Cardiovasc Electrophysiol. 1995;6:832–51.DOI
    29. 29.
      Grönefeld GC, Mauss O, Li YG, Klingenheben T, Hohnloser SH. Association between atrial fibrillation and appropriate implantable cardioverter defibrillator therapy: results from a prospective study. J Cardiovasc Electrophysiol. 2000;11:1208–14.DOI
    30. 30.
      Stein KM, Euler DE, Mehra R, Seidl K, Slotwiner DJ, Mittal S, et al. Do atrial tachyarrhythmias beget ventricular tachyarrhythmias in defibrillator recipients. J Am Coll Cardiol. 2002;40:335–40.DOI
    31. 31.
      Smit MD, Van Dessel PFHM, Rienstra M, Nieuwland W, Wiesfeld ACP, Tan ES, et al. Atrial fibrillation predicts appropriate shocks in primary prevention implantable cardioverter-defibrillator patients. Europace. 2006;8:566–72.DOI
    32. 32.
      Sakhuja R, Shah AJ, Keebler M, Thakur RK. Atrial fibrillation in patients with implantable defibrillators. Cardiol Clin. 2009;27:151–61.DOI
    33. 33.
      Petrac D, Radićc B, Radeljić V, Hamel D, Filipović J. Impact of atrioventricular node ablation and pacing therapy on clinical course in patients with permanent atrial fibrillation and unstable ventricular tachycardia induced by rapid ventricular response: follow-up study. Croat Med J. 2005 Dec;46(6):929-35.PubMed
    34. 34.
      Moss AJ, Greenberg H, Case RB, Zareba W, Hall J, Brown MW, et al. Long-term clinical course of patients after termination of ventricular tachyarrhythmia by an Implanted Defibrillator. Circulation. 2004;110:3760–5.DOI
    35. 35.
      Proietti R, Labos C, Davis M, Thanassoulis G, Santaganeli P, Russo V, et al. A systematic review and meta-analysis of the association between implantable cardioverter-defibrillator shocks and long-term mortality. Can J Cardiol. 2015;31:270–7.DOI
    36. 36.
      Mitchell LB, Pineda EA, Titus JL, Bartosch PM, Beneditt DG. Sudden death in patients with implantable cardioverter defibrillators. The importance of post-shock electromechanical dissociation. J Am Coll Cardiol. 2002;39:1323–8.DOI
    37. 37.
      Clementy N, Bodin A, Bisson A, Teixeira-Gomez AP, Roge S, Angouvlant D, et al. The defibrillation conundrum: new insights into the mechanisms of shock-related myocardial injury sustained from a life-saving therapy. Int J Mol Sci. 2021;22:1–17.DOI
    38. 38.
      Tereshchenko LG, Faddis MN, Fetics BJ, Zelik KE, Efimov IR, Berger RD. Transient local injury current in right ventricular electrogram after implantable cardioverter-defibrillator shock predicts heart failure progression. J Am Coll Cardiol. 2009;54:822–8.DOI
    39. 39.
      Tan NY, Roger VL, Killian JM, Cha YM, Noseworthy PA, Dunlay SM. Ventricular arrhythmias among patients with advanced heart failure: a population‐based study. J Am Heart Assoc. 2022;11:e023377.DOI
    40. 40.
      Santhanakrishnan R, Wang NA, Larsen MG, Mgnani JW, McManus DD, Lubitz SA, et al. Atrial fibrillation begets heart failure and vice versa: Temporal associations and differences in preserved versus reduced ejection fraction. Circulation. 2016;133:484–92.DOI
    41. 41.
      Aktaş MK, Younis A, Zareba W, Kutyifa V, Klein H, Daubert JP, et al. Survival after implantable cardioverter defibrillator shocks. J Am Coll Cardiol. 2021;77:2453–62.DOI
    42. 42.
      Tan VH, Wilton SB, Kuriachan V, Sumner GL, Exner DV. Impact of programming strategies aimed at reducing nonessential implantable cardioverter defibrillator therapies on mortality. A systematic review and meta-analysis. Circ Arrhythm Electrophysiol. 2014;7:164–70.DOI
    43. 43.
      Scott PA, Silberbauer J, McDonagh TA, Murgatroyd FD. Impact of prolonged implantable cardioverter-defibrillator arrhythmia detection times on outcomes: a meta-analysis. Heart Rhythm. 2014;11:828–35.DOI
    44. 44.
      Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA. 2002;288:3115–23.DOI
    45. 45.
      Barsheshet A, Moss AJ, McNitt S, Jons C, Glikson M, Klein HU, et al. Long-term implications of cumulative right ventricular pacing among patients with an implantable cardioverter-defibrillator. Heart Rhythm. 2011;8:212–8.DOI
    46. 46.
      Gasparini M, Proclemer A, Klersy C, Kloppe A, Ferrer JBM, Hersi A, et al. Effect of long-detection interval vs standard-detection interval for implantable cardioverter defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA. 2013;309:1903–11.DOI
    47. 47.
      Saeed M, Hanna I, Robotis D, Styperek R, Polosajian L, Khan A, et al. Programming implantable cardioverter-defibrillators in patients with primary prevention indication to prolong time to first shock: results from the PROVIDE study. J Cardiovasc Electrophysiol. 2014;25:52–9.DOI
    48. 48.
      Wilkoff BL, Fauchier L, Stiles MK, Morillo CA, Al-Khatib SM, Almendral J, et al. 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Europace. 2016;18:159–83.DOI
    49. 49.
      Wilkoff BL, Ousdigian KT, Sterns LD, Wang ZJ, Wilson RD, Morgan JM, et al. A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicentre EMPIRIC trial. J Am Coll Cardiol. 2006;48:330–9.DOI
    50. 50.
      Wathen MS, DeGroot PJ, Sweeney MO, Stark AJ, Otterness MF, Adkisson WO, et al. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. Circulation. 2004;110:2591–6.DOI
    51. 51.
      Brodine WN, Tung RT, Lee JK, Hockstad ES, Moss AJ, Zareba W, et al. Effects of Beta-Blockers on Implantable Cardioverter Defibrillator Therapy and Survival in the Patients with Ischemic Cardiomyopathy (from the Multicenter Automatic Defibrillator Implantation Trial-II). Am J Cardiol. 2005;96:691–5.DOI
    52. 52.
      Ruwald MH, Abu-Zeitone A, Jons C, Ruwald AC, McNitt S, Kutyifa V, et al. Impact of carvedilol and metoprolol on inappropriate implantable cardioverter-defibrillator therapy: the MADIT-CRT trial (Multicenter Automatic Defibrillator Implantation with Cardiac Resynchronization Therapy). J Am Coll Cardiol. 2013;62:1343–50.DOI
    53. 53.
      Metra M, Giubbini R, Nodari S, Boldi E, Moden MG, Dei CASL. Differential Effects of beta-blockers in patients with heart failure. A prospective, randomized, double-blind comparison of the long-term effects of metoprolol versus carvedilol. Circulation. 2000;102:546–51.DOI
    54. 54.
      Connolly SJ, Dorian P, Roberts RS, Gent M, Bailin S, Fain ES, et al. Comparison of beta-blockers, amiodarone plus beta-blockers, or sotalol for prevention of shocks from implantable cardioverter defibrillators: the OPTIC Study: a randomized trial. JAMA. 2006;295:165–71.DOI
    55. 55.
      Di Biase L, Mohanty P, Mohanty S, Santangeli P, Trivedi C, Lakkireddy D, et al. Ablation versus amiodarone for treatment of persistent atrial fibrillation in patients with congestive heart failure and an implanted device. Results from the AATAC multicenter randomized trial. Circulation. 2016;133:1637–44.DOI
    56. 56.
      Kosiuk J, Nedios S, Darma A, Rolf S, Richter S, Arya A, et al. Impact of single atrial fibrillation catheter ablation on implantable cardioverter defibrillator therapies in patients with ischaemic and non-ischaemic cardiomyopathies. Europace. 2014;16:1322–6.DOI
    57. 57.
      Miyazaki S, Taniguchi H, Kusa S, Komatsu Y, Ichihara N, Takagi T, et al. Catheter ablation of atrial tachyarrhythmias causing inappropriate implantable cardioverter-defibrillator shocks. Europace. 2015;17:289–94.DOI
    58. 58.
      Gasparini M, Kloppe A, Lunati M, Anselme F, Landolina M, Martinez-Ferrer JB, et al. Atrioventricular junction ablation in patients with atrial fibrillation treated with cardiac resynchronization therapy: positive impact on ventricular arrhythmias, implantable cardioverter-defibrillator therapies and hospitalizations: atrioventricular junction ablation in CRT patients with AF. Eur J Heart Fail. 2018;20:1472–81.DOI
    59. 59.
      Dickstein K, Normand C, Auricchio A, Bogale N, Cleland JG, Gitt A, et al. CRT Survey II: a European Society of Cardiology survey of cardiac resynchronisation therapy in 11088 patients - who is doing what to whom and how? Eur J Heart Fail. 2018;20:1039–51.DOI