Dyspnea after cardiac surgery — two unexpected echocardiographic solutions to a single symptom

    Authors

    Keywords

    echocardiography, cardiac surgery, valve dehiscence

    DOI

    https://doi.org/10.15836/ccar.2015.70

    Full Text

    ## Introduction We present two cases of the extremely rare and life-threatening cardiac surgery complications in patients presenting with dyspnea as the leading complaint, with echocardiographic imaging leading to a prompt and precise diagnosis. ## Case 1 A 41-year-old man was admitted to our department with the history of progressive dyspnea on exertion. Two years earlier a composite mechanical-valved conduit aortic root replacement (Bentall operation) had been performed due to severe regurgitation of the bicuspid aortic valve coupled with the aneurysm of the ascending aorta. Transthoracic echocardiogram demonstrated a 10 cm large aortic pseudo-aneurysm with the total dehiscence of the valved conduit at the site of the proximal anastomosis (Figure 1). The conduit was floating in the pseudo-aneurysmal cavity and was kept loosely in place by the tension of the coronary arteries, with preserved flow through the mechanical valve. Emergency surgery was performed. Intraoperatively, no signs of endocarditis were found, leaving suture line tension and connective tissue quality issue (due to underlying disease, namely BAV) as possible risk factors associated with the occurence of the complication. Figure 1. Parasternal long-axis view demonstrating the total separation of the mechanical-valved conduit (large arrow) within large pseudoaneurysm of the ascending aorta (delineated by the small arrrows) from the left ventricular outflow tract. ## Case 2 A 63-year-old man was referred to our department due to marked progressive dyspnea. A complex surgical procedure, namely aortic valve replacement with stentless bioprosthesis, ascending aorta and aortic arch reconstruction with dacron graft, triple aortocoronary bypass and mitral valve annuloplasty had been performed two months earlier in an out-of-country surgical center. Transthoracic and 3D transesophageal ecocardiography revealed total dehiscence and migration of the mitral annuloplasty ring to the left ventricular outflow tract through the perforation of the anterior mitral leaflet, approximately 1 cm from the aorticomitral junction, causing severe mitral regurgitation (Figure 2). The patient recovered uneventfully after successful mitral valve replacement (with mechanical prosthesis). Figure 2. 3D transesophageal echocardiogram demonstrating perforation of the anterior mitral leaflet by migrating mitral annuloplasty ring (arrows), from the left atrial view. ## Conclusion Due to its ready availability and high accuracy, echocardiography plays pivotal role in early diagnosis and appropriate management of the various complications following cardiac surgery.

    Cardiologia Croatica
    Back to search

    Dyspnea after cardiac surgery — two unexpected echocardiographic solutions to a single symptom

    Abstract
    Issue3-4
    Published
    Pages70
    PDF via DOIhttps://doi.org/10.15836/ccar.2015.70
    echocardiography
    cardiac surgery
    valve dehiscence

    Authors

    Zeljko Baricevic*ORCIDUniversity of Zagreb School of Medicine, University Hospital Centre Zagreb, Zagreb, Croatia
    Dejan DosenORCIDUniversity of Zagreb School of Medicine, University Hospital Centre Zagreb, Zagreb, Croatia
    Sandra VecericORCIDUniversity of Zagreb School of Medicine, University Hospital Centre Zagreb, Zagreb, Croatia
    Ivica SafradinORCIDUniversity of Zagreb School of Medicine, University Hospital Centre Zagreb, Zagreb, Croatia
    Darko AnicORCIDUniversity of Zagreb School of Medicine, University Hospital Centre Zagreb, Zagreb, Croatia
    Jadranka Separovic HanzevackiORCIDUniversity of Zagreb School of Medicine, University Hospital Centre Zagreb, Zagreb, Croatia

    *Correspondence email: zbaricev@gmail.com

    Full Text

    Introduction

    We present two cases of the extremely rare and life-threatening cardiac surgery complications in patients presenting with dyspnea as the leading complaint, with echocardiographic imaging leading to a prompt and precise diagnosis.

    Case 1

    A 41-year-old man was admitted to our department with the history of progressive dyspnea on exertion. Two years earlier a composite mechanical-valved conduit aortic root replacement (Bentall operation) had been performed due to severe regurgitation of the bicuspid aortic valve coupled with the aneurysm of the ascending aorta. Transthoracic echocardiogram demonstrated a 10 cm large aortic pseudo-aneurysm with the total dehiscence of the valved conduit at the site of the proximal anastomosis (Figure 1). The conduit was floating in the pseudo-aneurysmal cavity and was kept loosely in place by the tension of the coronary arteries, with preserved flow through the mechanical valve. Emergency surgery was performed. Intraoperatively, no signs of endocarditis were found, leaving suture line tension and connective tissue quality issue (due to underlying disease, namely BAV) as possible risk factors associated with the occurence of the complication.

    Figure 1. Parasternal long-axis view demonstrating the total separation of the mechanical-valved conduit (large arrow) within large pseudoaneurysm of the ascending aorta (delineated by the small arrrows) from the left ventricular outflow tract.

    Case 2

    A 63-year-old man was referred to our department due to marked progressive dyspnea. A complex surgical procedure, namely aortic valve replacement with stentless bioprosthesis, ascending aorta and aortic arch reconstruction with dacron graft, triple aortocoronary bypass and mitral valve annuloplasty had been performed two months earlier in an out-of-country surgical center. Transthoracic and 3D transesophageal ecocardiography revealed total dehiscence and migration of the mitral annuloplasty ring to the left ventricular outflow tract through the perforation of the anterior mitral leaflet, approximately 1 cm from the aorticomitral junction, causing severe mitral regurgitation (Figure 2). The patient recovered uneventfully after successful mitral valve replacement (with mechanical prosthesis).

    Figure 2. 3D transesophageal echocardiogram demonstrating perforation of the anterior mitral leaflet by migrating mitral annuloplasty ring (arrows), from the left atrial view.

    Conclusion

    Due to its ready availability and high accuracy, echocardiography plays pivotal role in early diagnosis and appropriate management of the various complications following cardiac surgery.